2-Pack Analyzer System
B
–
26
4900002256 rev. B 4-13-21
Disposal of Used Scrubbers
1.
Discard used scrubber and scrubber indicator in an appropriate leak-
proof receptacle.
Instrument Issues
If the instrument does not appear to be hampered by gas leaks, contamination,
excessive sampling gas temperatures and pressures, or electrical noise, refer
to Table B–1 on page B–26 before contacting Service. Refer to
on page B-31 if assistance is required.
Depleted H
2
S scrubbers and scrubber indicators contain
predominantly Copper (II) Sulfide [CAS# 1317-40-4] with some
remaining Copper (II) Oxide [CAS# 1317-38-0] and basic cupric
carbonate [CAS# 12069-69-1], each of which are odorless dark
powders that require few special precautions other than avoiding
contact with the internal substances, keeping the scrubber tightly
sealed and protecting the contents against humidity.
Table B–1
Potential analyzer issues and solutions
Symptom
Response
Non-Operation (at start up)
Is the power connected to both the
analyzer and power source? Is the
switch on?
Non-Operation (after start up)
Is the power source good? (100 to
250 VAC at 50-60 Hz, 9 to 16 VDC,
18 to 32 VDC).
Check fuse(s). If bad, replace with
equivalent amperage, slow-blow fuse.
Refer to Table 1–1 on page 1–15 or
B-31 for service information.
Laser Power Low Alrm
Turn off the power to the unit and
check the optical head cables for a
loose connection.
Do not disconnect
or reconnect any optical head
cables with the power connected.
Check the inlet and outlet tubes to see
if they are under any stress. Remove
the connections to the inlet and outlet
tubes and see if the power goes up.
Perhaps the existing tubing needs to be
replaced with stainless steel flexible
tubing.
Summary of Contents for SS1000
Page 2: ......
Page 4: ......
Page 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Page 121: ...www spectrasensors com contact 4900002256 Rev B ...