2-Pack Analyzer System
3
–
12
4900002256 rev. B 4-13-21
The
Concentra High Alarm
(Analyzer A),
Assignable Alarm
(Analyzer B)
and
General Fault Alarm
trigger SPDT relays located inside the respective
analyzer electronics enclosures.
The relays for the
Concentra High Alarm
and
Assignable Alarm
are
configured to be normally de-energized with the dry contact changing state on
alarm. The
Concentra High Alarm
and
Assignable Alarm
outputs can be
wired for OPEN or CLOSED depending on which terminals are used normally
open (NO) or normally closed (NC).
The relay for the
General Fault Alarm
for each analyzer is configured to be
fail-safe (or normally energized) so the dry contact will open in the event of
power loss or
General Fault Alarm
. Thus, the cables for each
General Fault
Alarm
should be wired to the common and NO terminals.
Connections can be made with customer supplied cables for the current loops
and alarms and factory-supplied cables for the serial connections. Refer to the
wiring diagram in Appendix A.
To connect Analyzer A signal and alarm cables
1.
Disconnect power from analyzer and open the Analyzer A electronics
enclosure cover. Take care not to disturb the electrical assembly
inside.
2.
Run conduit from the signal/alarm receiving station to the conduit
hub in the upper left-hand corner of the Analyzer A electronics
enclosure labeled for signal input/output. Refer to
on page 3-10 and analyzer drawings in
3.
Pull the customer supplied cables for the current loop, alarms, digital
input, Ethernet and the serial cable through the conduit into the
Analyzer A electronics enclosure. (No connections are required for
the analyzer to run and function properly. Only make connections for
signals that will be used for communication with the analyzer).
Hazardous voltage and risk of electric shock
. Be sure power
to the analyzer is turned off before opening either electronics
enclosure and making any connections.
Hazardous voltage and risk of electric shock
. Be sure power
to the analyzer is turned off before opening the electronics
enclosure and making any connections.
Conduit seals should be used where appropriate in compliance
with local regulations.
Summary of Contents for SS1000
Page 2: ......
Page 4: ......
Page 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Page 121: ...www spectrasensors com contact 4900002256 Rev B ...