2-Pack Analyzer System
3
–
18
4900002256 rev. B 4-13-21
To calibrate the analog output (Analyzer A)
1.
Connect a calibrator and digital multi-meter into the circuit.
2.
On the analyzer keypad, press
#2
(
Mode 2
), the password (3142)
and
*
.
3.
Continue pressing the
*
key until the
4-20 mA Test
parameter
displays.
4.
Enter the desired percentage of full scale and press
*
.
a.
Set 4-20 mA Test = 0; this displays the 4 mA on the AO circuit
when
#5
(
Mode 5
) is pressed.
b.
Set 4-20 mA Test = 50; this displays the 12 mA on the AO circuit
when
#5
is pressed.
c.
Set 4-20 mA Test = 100; this displays the 20 mA on the AO
circuit when
#5
is pressed.
5.
Press
#
and
1
to return to Normal Mode.
To calibrate the analog output (Analyzer B)
1.
Connect a calibrator and digital multi-meter into the circuit.
2.
On the analyzer keypad, press
#2
(
Mode 2
), the password (3142)
and
*
.
3.
Continue pressing the
*
key until the
4-20 mA % Test
parameter
displays.
4.
Set the current loop output to 20 mA by setting to 100.
5.
Adjust the receiver calibration control to read the appropriate value.
the current loop output of 20 mA corresponds to the concentration
value set by the
20 mA Value
parameter.
<SET PARAMETER MODE>
AO 4-20 mA Test
0.00000
Enter a value (%)
<SET PARAMETER MODE>
4-20 mA % Test
101
Enter a % (101=Off)
Summary of Contents for SS1000
Page 2: ......
Page 4: ......
Page 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Page 121: ...www spectrasensors com contact 4900002256 Rev B ...