Installation
Hardware Installation and Maintenance Manual
3
–
17
To change 4‐20 mA board from source to sink (Analyzer A or B)
1.
Disconnect power from the analyzer and open the electronics
enclosure cover. Take care not to disturb the electrical assembly
inside.
2.
Locate the relay control board in the upper right of the electronics
enclosure, as shown in Figure 1–6 on page 1–11 or Figure 1–7 on
page 1–12 (Analyzer A) or Figure 1–8 on page 1–13 or Figure 1–9
on page 1–14 (Analyzer B).
3.
Remove the jumper (JMP1), shown in Figure 3–8 on page 3–17,
connecting the center pin to point “A.”
4.
For 4-20 mA sink, carefully replace the jumper to connect the center
hole with point “P.”
5.
Reconnect power to the analyzer. Confirm the 4 mA (minimum) and
20 mA (maximum) points.
6.
Close and tighten the analyzer electronics enclosure cover.
Needle nose pliers may be required to remove the jumper
.
For Analyzer A, s
“To calibrate the analog output (Analyzer
. For the
Analyzer
B
, refer to
Figure 3–8
Analyzer 4-20 mA board
JMP1
POTENTIOMETERS
Summary of Contents for SS1000
Page 2: ......
Page 4: ......
Page 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Page 121: ...www spectrasensors com contact 4900002256 Rev B ...