Proline Promass S 100
Hauser
67
Reliable operation
• Operation in the following languages:
– Via "FieldCare" operating tool:
English, German, French, Spanish, Italian, Chinese, Japanese
– Via Web browser:
English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese,
Japanese, Bahasa (Indonesian), Vietnamese, Czech
• Uniform operating philosophy applied to operating tools and Web browser
• If replacing the electronic module, transfer the device configuration via the plug-in memory
(HistoROM DAT) which contains the process and measuring device data and the event logbook. No
need to reconfigure.
For devices with Modbus RS485, the data recovery function is implemented without the plug-in
memory (HistoROM DAT).
Efficient diagnostics increase measurement availability
• Troubleshooting measures can be called up via the operating tools and Web browser
• Diverse simulation options
• Status indicated by several light emitting diodes (LEDs) on the electronic module in the housing
compartment
Local display
The local display is only available with the following device version:
Order code for "Display; Operation", option
B
: 4-line; via communication
Display element
• 4-line liquid crystal display with 16 characters per line.
• White background lighting; switches to red in event of device errors.
• Format for displaying measured variables and status variables can be individually configured.
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F). The readability of
the display may be impaired at temperatures outside the temperature range.
Remote operation
Via HART protocol
This communication interface is present in the following device version:
Order code for "Output", option
B
: 4-20 mA HART, pulse/frequency/switch output
1
2
3
5
7
4
6
A0016948
47
Options for remote operation via HART protocol
1
Control system (e.g. PLC)
2
Field Communicator 475
3
Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM)
4
Commubox FXA195 (USB)
5
Field Xpert SFX350 or SFX370
6
VIATOR Bluetooth modem with connecting cable
7
Transmitter
Via Ethernet-based fieldbus
This communication interface is present in the following device version:
Order code for "Output", option
N
: EtherNet/IP