Output
Output signal
Current output
Current output
4-20 mA HART (active)
Maximum output values
• DC 24 V (no flow)
• 22.5 mA
Load
0 to 700 Ω
Resolution
0.38 µA
Damping
Adjustable: 0.07 to 999 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
Function
Can be set to pulse, frequency or switch output
Version
Passive, open collector
Maximum input values
• DC 30 V
• 25 mA
Voltage drop
For 25 mA: ≤ DC 2 V
Pulse output
Pulse width
Adjustable: 0.05 to 2
000 ms
Maximum pulse rate
10
000 Impulse/s
Pulse value
Adjustable
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
Frequency output
Output frequency
Adjustable: 0 to 10
000 Hz
Damping
Adjustable: 0 to 999 s
Pulse/pause ratio
1:1
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Switch output
Switching behavior
Binary, conductive or non-conductive
Switching delay
Adjustable: 0 to 100 s