Proline Promass S 100
12
Hauser
Web browser
Plain text display
With information on cause and remedial measures
Light emitting diodes (LED)
Status information
Status indicated by various light emitting diodes
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
• EtherNet/IP network available
• EtherNet/IP connection established
Ex connection data
These values only apply for the following device version:
Order code for "Output", option
M
: Modbus RS485, for use in intrinsically safe areas
Safety Barrier Promass 100
Safety-related values
Terminal numbers
Supply voltage
Signal transmission
2 (L-)
1 (L+)
26 (A)
27 (B)
U
nom
= DC 24 V
U
max
= AC 260 V
U
nom
= DC 5 V
U
max
= AC 260 V
Intrinsically safe values
Terminal numbers
Supply voltage
Signal transmission
20 (L-)
10 (L+)
62 (A)
72 (B)
U
o
= 16.24 V
I
o
= 623 mA
P
o
= 2.45 W
For IIC
*
: L
o
= 92.8 µH, C
o
= 0.433 μF, L
o
/R
o
= 14.6 μH/Ω
For IIB
*
: L
o
= 372 µH, C
o
= 2.57 μF, L
o
/R
o
= 58.3 μH/Ω
*
The gas group depends on the sensor and nominal diameter.
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
Transmitter
Intrinsically safe values
Order code for
"Approvals"
Terminal numbers
Supply voltage
Signal transmission
20 (L-)
10 (L+)
62 (A)
72 (B)
• Option
BM
: ATEX II2G + IECEx Z1 Ex ia, II2D Ex tb
• Option
BO
: ATEX II1/2G + IECEx Z0/Z1 Ex ia, II2D
• Option
BQ
: ATEX II1/2G + IECEx Z0/Z1 Ex ia
• Option
BU
: ATEX II2G + IECEx Z1 Ex ia
• Option
C2
: CSA C/US IS Cl. I, II, III Div. 1
• Option
85
: ATEX II2G + IECEx Z1 Ex ia + CSA C/US
IS Cl. I, II, III Div. 1
U
i
= 16.24 V
I
i
= 623 mA
P
i
= 2.45 W
L
i
= 0 µH
C
i
= 6 nF
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
Low flow cut off
The switch points for low flow cut off are user-selectable.