Technical data
Proline Promass E 100 Modbus RS485
92
Hauser
Shock resistance
Shock, half-sine according to IEC 60068-2-27
6 ms 30 g
Shock resistance
Shock due to rough handling following IEC 60068-2-31
Interior cleaning
• SIP cleaning
• CIP cleaning
Electromagnetic
compatibility (EMC)
• As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
• Complies with emission limits for industry as per EN 55011 (Class A)
Details are provided in the Declaration of Conformity.
16.9 Process
Medium temperature range
–40 to +150 °C (–40 to +302 °F)
Dependency of ambient temperature on medium temperature
T
a
T
m
B
A
A0031121
21
Exemplary representation, values in the table below.
T
a
Ambient temperature range
T
m
Medium temperature
A
Maximum permitted medium temperature T
m
at T
a max
= 60 °C (140 °F); higher medium temperatures T
m
require a reduced ambient temperature T
a
B
Maximum permitted ambient temperature T
a
for the maximum specified medium temperature T
m
of the
sensor
Values for devices used in the hazardous area:
Separate Ex documentation (XA) for the device .
Density
0 to 5
000 kg/m
3
(0 to 312 lb/cf)
Pressure-temperature
ratings
An overview of the pressure-temperature ratings for the process connections is
provided in the "Technical Information" document