Proline Promass E 100 Modbus RS485
Commissioning
Hauser
49
10
Commissioning
10.1 Function check
Before commissioning the device, make sure that the post-installation and post-
connection checks have been performed.
• "Post-mounting check" checklist → 23
• "Post-connection check" checklist → 34
10.2 Establishing a connection via FieldCare
• For FieldCare connection
• For establishing a connection via FieldCare → 42
• For FieldCare user interface → 43
10.3 Configuring the measuring device
The
Setup
menu with its submenus contains all parameters needed for standard
operation.
Structure of the "Setup" menu
Setup
→
System units
Select medium
Communication
Low flow cut off
Partially filled pipe detection
10.3.1 Setting the system units
In the
System units
submenu, you can configure the units of all measured values.
Navigation path
"Setup" menu → Advanced setup → System units
Structure of the submenu
System units
→
Mass flow unit
Mass unit
Volume flow unit
Volume unit
Corrected volume flow unit
Corrected volume unit
Density unit
Reference density unit