Proline Promass E 100 Modbus RS485
Technical data
Hauser
89
Base repeatability
Mass flow and volume flow (liquids)
±0.075 % o.r.
±0.05 % o.r. (calibration option, for mass flow)
Mass flow (gases)
±0.35 % o.r.
Density (liquids)
±0.00025 g/cm
3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
Response time
The response time depends on the configuration (damping).
Influence of medium
temperature
Mass flow and volume flow
o.f.s. = of full scale value
When there is a difference between the temperature for zero point adjustment and the
process temperature, the additional measured error of the sensor is typically ±0.0002 %
o.f.s./°C (±0.0001 % o. f.s./°F).
The effect is reduced if zero point adjustment is performed at process temperature.
Density
When there is a difference between the density calibration temperature and the process
temperature, the typical measured error of the sensor is
±0.0001 g/cm
3
/°C (±0.00005 g/cm
3
/°F). Field density calibration is possible.
[kg/m ]
3
[°C]
-40
0
-80
40 80 120 160 200 240 280 320 [°F]
-40
0
50
100
150
0
2
4
6
8
10
14
12
A0016609
20
Field density calibration, for example at +20 °C (+68 °F)
Temperature
±0.005 · T °C (± 0.005 · (T – 32) °F)
Influence of medium
pressure
The table below shows the effect on accuracy of mass flow due to a difference between
calibration pressure and process pressure.
Summary of Contents for Proline Promass E 100
Page 107: ......