Hauser
Flow characteristic values for Promag H sensor (SI units)
Flow characteristic values for Promag H sensor (US units)
Nominal
diameter
Recommended flow rate
Factory settings
[mm]
[inch]
Min./max. full scale value
(v
≈
0.3 or 10 m/s)
Full scale value
(v
≈
2.5 m/s)
Pulse value
(
≈
2 pulse/s)
Low flow cut off
(v
≈
0.04 m/s)
2
1/12"
0.06 to 1.8 dm
3
/min
0.5 dm
3
/min
0.005 dm
3
0.01 dm
3
/min
4
5/32"
0.25 to 7 dm
3
/min
2 dm
3
/min
0.025 dm
3
0.05 dm
3
/min
8
5/16"
1 to 30 dm
3
/min
8 dm
3
/min
0.10 dm
3
0.1 dm
3
/min
15
1/2"
4 to 100 dm
3
/min
25 dm
3
/min
0.20 dm
3
0.5 dm
3
/min
25
1"
9 to 300 dm
3
/min
75 dm
3
/min
0.50 dm
3
1 dm
3
/min
32
1 1/4"
15 to 500 dm
3
/min
125 dm
3
/min
1.00 dm
3
2 dm
3
/min
40
1 1/2"
25 to 700 dm
3
/min
200 dm
3
/min
1.50 dm
3
3 dm
3
/min
50
2"
35 to 1100 dm
3
/min
300 dm
3
/min
2.50 dm
3
5 dm
3
/min
65
2 1/2"
60 to 2000 dm
3
/min
500 dm
3
/min
5.00 dm
3
8 dm
3
/min
80
3"
90 to 3000 dm
3
/min
750 dm
3
/min
5.00 dm
3
12 dm
3
/min
100
4"
145 to 4700 dm
3
/min
1200 dm
3
/min
10.00 dm
3
20 dm
3
/min
Nominal
diameter
Recommended flow rate
Factory settings
[inch]
[mm]
Min./max. full scale value
(v
≈
0.3 or 10 m/s)
Full scale value
(v
≈
2.5 m/s)
Pulse value
(
≈
2 pulse/s)
Low flow cut off
(v
≈
0.04 m/s)
1/12"
2
0.015 to 0.5 gal/min
0.1 gal/min
0.001 gal
0.002 gal/min
5/32"
4
0.07 to 2 gal/min
0.5 gal/min
0.005 gal
0.008 gal/min
5/16"
8
0.25 to 8 gal/min
2 gal/min
0.02 gal
0.025 gal/min
1/2"
15
1.0 to 27 gal/min
6 gal/min
0.05 gal
0.10 gal/min
1"
25
2.5 to 80 gal/min
18 gal/min
0.20 gal
0.25 gal/min
1 1/4"
32
4 to 130 gal/min
30 gal/min
0.20 gal
0.5 gal/min
1 1/2"
40
7 to 190 gal/min
50 gal/min
0.50 gal
0.75 gal/min
2"
50
10 to 300 gal/min
75 gal/min
0.50 gal
1.25 gal/min
2 1/2"
65
16 to 500 gal/min
130 gal/min
1 gal
2.0 gal/min
3"
80
24 to 800 gal/min
200 gal/min
2 gal
2.5 gal/min
4"
100
40 to 1250 gal/min
300 gal/min
2 gal
4.0 gal/min