Proline Promag 55 FOUNDATION Fieldbus
Troubleshooting
Hauser
87
041
Device status message (FF):
T-DAT failure –
Err. No. 041
Local display:
S: TRANSM. HW-DAT
$
: # 041
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Cause
:
1.
T DAT is not correctly plugged into the
amplifier board (or is missing).
2.
T-DAT is defective.
Remedy:
1.
Check whether the T-DAT is correctly plugged
into the amplifier board.
2.
Replace the T-DAT if it is defective.
Spare parts
Check whether the new replacement DAT is
compatible with the existing electronics.
Check the:
- Spare part set number
- Hardware revision code
3.
Replace measuring electronics boards if
necessary
Spare parts
4.
Plug the T-DAT into the amplifier board.
Transducer_Error = Electronics
failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
042
Device status message (FF):
T-DAT data inconsistent –
Err. No. 042
Local display:
S: TRANSM. SW-DAT
$
: # 042
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Transducer_Error = Data
integrity error
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
No. # 1xx
→
Software error
101
Device status message (FF):
GAIN ERROR AMPLIFIER –
Err. No. 101
Local display:
S: GAIN ERROR AMP.
$
: # 101
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Cause:
Gain deviation compared to reference gain is
greater than 2%.
Remedy:
Replace the amplifier board
Transducer_Error = Electronics
failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
121
Device status message (FF):
Software compatibility
problem amplifier –
I/O module –
Err. No. 121
Local display:
S: A/C COMPATIB.
!: # 121
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Cause:
Due to different software versions, I/O board and
amplifier board are only partially compatible
(possibly restricted functionality).
!
Note!
– Appears on the display as a notice message for
only 30 seconds (with listing in "Previous system
condition" function).
– This occurrence of different software versions
can occur if only one electronics board has been
replaced; the extended software functionality is
not available. The previously existing software
functionality is still working and the
measurement possible.
Remedy:
Module with lower software version must either be
updated by "FieldCare" using the required
(recommended) software version, or the module
must be replaced.
Transducer_Error = I/O failure
(input/output error)
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
No.
Error messages:
FOUNDATION Fieldbus
(FF)*
(local display)
Transducer Block
error messages
Analog Input function block
error messages
Cause/remedy