PROline Promag 23
9 Trouble-shooting
Hauser
63
S
$
SENSOR SW DAT
# 032
Sensor:
Error accessing the calibration
values stored in the S-DAT.
1.
Check whether the S-DAT is
correctly plugged into the
amplifier board
2.
Replace the S-DAT if it is
defective. Spare parts
Before replacing the DAT,
check that the new, replace-
ment DAT is compatible with
the measuring electronics.
Check the:
– Spare part set number
– Hardware revision code
3.
Replace measuring electro-
nics boards if necessary.
Spare parts
S
$
TRANSM. SW-DAT
# 042
Transmitter DAT:
Error accessing the calibration
values stored in the T-DAT.
1.
Check whether the T-DAT is
correctly plugged into the I/O
board
2.
Replace the T-DAT if it is
defective.
Spare parts
Before replacing the DAT,
check that the new, replace-
ment DAT is compatible with
the measuring electronics.
Check the:
– Spare part set number
– Hardware revision code
3.
Replace measuring electro-
nics boards if necessary.
Spare parts
S
$
A / C COMPATIB.
# 051
The I/O board and the amplifier
board are not compatible.
Use only compatible modules and
boards. Check the compatibility of
the modules used.
Check the:
– Spare part set number
– Hardware revision code
Nr. # 1xx
Software error
S
$
CHECKSUM TOTAL
# 111
Totalizer checksum error
1.
Restart the measuring device
2.
Replace the amplifier board if
necessary.
Spare parts
Type
Error message / No.
Cause
Remedy / spare part
Summary of Contents for PROline promag 23
Page 4: ...Contents PROline Promag 23 4 Endress Hauser ...
Page 10: ...2 Identification PROline Promag 23 10 Endress Hauser ...
Page 30: ...3 Installation PROline Promag 23 30 Endress Hauser ...
Page 40: ...4 Wiring PROline Promag 23 40 Endress Hauser ...
Page 52: ...5 Operation PROline Promag 23 52 Endress Hauser ...
Page 56: ...6 Commissioning PROline Promag 23 56 Endress Hauser ...
Page 58: ...7 Maintenance PROline Promag 23 58 Endress Hauser ...
Page 102: ...10 Technical data PROline Promag 23 102 Endress Hauser ...