PROline Promag 10
5 Operation
Hauser
59
5.4.5
Device status / error messages
You can read the extended device status, in this case, current error messages, via
command "48". The command delivers bit-encoded information (see table below).
!
Note!
•
Detailed information on the device status messages and error messages, and how
they are rectified, can be found on Page 67 ff.!
•
Bits not listed are not assigned.
Byte
Bit
Error no.
Short error description (
0
0
001
Serious device error.
1
011
Faulty amplifier EEPROM.
2
012
Error when accessing data of the amplifier EEPROM.
3
3
111
Totalizer checksum error
5
0
321
Coil current of sensor is outside tolerance.
7
3
351
Current output: the current flow is outside the set range.
8
3
359
Pulse output: the pulse output frequency is outside the set range.
10
7
401
Measuring tube partially filled or empty
11
2
461
EPD adjustment is not possible as the fluid conductivity is too low
or too high.
4
463
The EPD adjustment values for empty pipe or full pipe are
identical, i.e. faulty.
12
7
501
Loading new measuring amplifier software version. No other
commands possible at this point.
14
3
601
Positive zero return active.
18
3
691
Simulation of failsafe mode (outputs) active
4
692
Simulation volume flow