Calibration with partially
filled vessel
If the following calibration is initially carried out with a hydrostatic pressure sensor using
water (
ρ
= 1.0), density can be measured and read off from V8H7.
•
If the product changes, the system can be quickly adapted to the new
product by entering a its density at V8H7.
This ensures that the correct level is
measured as the vessel is filled
.
Sensor calibration
Liquiphant limit
level probe
#
Matrix
Entry
Remarks
1
V9H5
Reset, enter probe
V3H5/V3H6
constants, page 23
2
V8H0
5
Calibration correction
3
-
»E«
Register entry
4
V8H2
e.g. 2 s
Switching delay
5
-
»E«
Register entry
6
V8H3
e.g. 75% Installation height
7
-
»E«
Register entry
8
V8H4
0
Sensor = Liquiphant
9
-
»E«
Register entry
Level calibration
#
Matrix
Entry
Remarks
10 V8H7
1
At »empty« level,
set corr. factor to 1
11 -
»E«
Register entry
12 V0H1
e.g. 10% »Empty« level
13 -
»E«
Register entry
14 V8H7
1
At »full« level
set corr. factor to 1
15 -
»E«
Register entry
16 V0H2
e.g. 90% »Full« level
17 -
»E«
Register entry
Note!
•
The best accuracy is attained when the
difference between »full« and »empty«
levels is as great as possible.
•
If the operating mode is temporarily
quit, the Prolevel measures by using
the sensitivity V3H2 x V8H3. The
indication remains correct as long as
the same product is in the tank.
Next step
…
•
If appropriate, linearisation, p. 25
…
27
•
Analogue output and relays, p. 30
…
33
— assign relays 2a/2b to channel 1.
100%
in
stal
la
ti
on
hei
ght
m
easur
ing r
ange
e.g. 75%
»Empty« e.g. 10%
0%
Note!
V0H1, V8H7
V0H2, V8H7
BA142_46
»Full« e.g. 90%
0%
m
easur
ing r
ange
Fig. 5.4:
Parameters for calibration with partially filled vessel
Chapter 5: Level Measurement with Limit Switch
Prolevel FMC 661
38