Orbisint CPS11D and CPS11
Hauser
9
Reference system
Version AA, BA, FA: Ag/AgCl reference lead with Advanced Gel 3M KCl, AgCl-free
Version AS:
Ag/AgCl reference lead with Advanced Gel, saturated KCl (> 3M KCl) with salt
rings, AgCl-free
The following are indicative of used salt rings (fixed supply of KCl) under
constant process conditions (e.g. stable temperature and flow):
• a continuous upward trend in the pH value (to alkaline pH values)
• a continuous downward trend in the zero point (to acidic pH values)
following adjustment during calibration
BT version:
Ag/AgCl reference lead with ion trap and Advanced Gel 3M KCl
Certificates and approvals
Ex approval for CPS11D
• ATEX II 1G Ex ia IIC T3/T4/T6 Ga
• FM / CSA Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters
Hazardous area versions of the digital sensors with Memosens technology are marked by a red-
orange ring in the plug-in head.
Ex approval for CPS11
(TOP68)
• ATEX II 1G Ex ia IIC T3/T4/T6 Ga
• FM Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters
Biocompatibility
Cytotoxicity verified in accordance with:
USP 2009, chapter <88> (USP Class VI) for diaphragm
TÜV certificate for
Memosens plug-in head
Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure
TÜV certificate for TOP68
plug-in head
Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure
CPS11D electromagnetic
compatibility
Interference emission and interference immunity as per EN 61326: 2012
Ordering information
Product page
www.endress.com/cps11d
www.endress.com/cps11
Product Configurator
The navigation area is located on the right of the product page.
1. Under "Device support" click "Configure your selected product".
The Configurator opens in a separate window.
2. Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.
3. Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of
the screen.
Scope of delivery
The scope of delivery includes:
• Sensor in the version ordered
• Technical Information