background image

Orbisint CPS11D and CPS11

Hauser

9

Reference system

Version AA, BA, FA: Ag/AgCl reference lead with Advanced Gel 3M KCl, AgCl-free
Version AS:

Ag/AgCl reference lead with Advanced Gel, saturated KCl (> 3M KCl) with salt

rings, AgCl-free

The following are indicative of used salt rings (fixed supply of KCl) under

constant process conditions (e.g. stable temperature and flow):

• a continuous upward trend in the pH value (to alkaline pH values)

• a continuous downward trend in the zero point (to acidic pH values)

following adjustment during calibration

BT version:

Ag/AgCl reference lead with ion trap and Advanced Gel 3M KCl

Certificates and approvals

Ex approval for CPS11D

• ATEX II 1G Ex ia IIC T3/T4/T6 Ga

• FM / CSA Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters

Hazardous area versions of the digital sensors with Memosens technology are marked by a red-

orange ring in the plug-in head.

Ex approval for CPS11

(TOP68)

• ATEX II 1G Ex ia IIC T3/T4/T6 Ga

• FM Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters

Biocompatibility

Cytotoxicity verified in accordance with:

USP 2009, chapter <88> (USP Class VI) for diaphragm

TÜV certificate for

Memosens plug-in head

Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure

TÜV certificate for TOP68

plug-in head

Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure

CPS11D electromagnetic

compatibility

Interference emission and interference immunity as per EN 61326: 2012

Ordering information

Product page

www.endress.com/cps11d
www.endress.com/cps11

Product Configurator

The navigation area is located on the right of the product page.

1. Under "Device support" click "Configure your selected product".

 The Configurator opens in a separate window.

2. Select all the options to configure the device in line with your requirements.

 In this way, you receive a valid and complete order code for the device.

3. Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of

the screen.

Scope of delivery

The scope of delivery includes:

• Sensor in the version ordered

• Technical Information

Summary of Contents for Orbisint CPS11

Page 1: ...PTFE ring junction Can be used at pressures up to 17 bar abs 246 psi Process glass also for very alkaline applications BA and BT versions Process glass for applications in media containing hydrofluori...

Page 2: ...tegrated Ag AgCl reference system serves as the required reference electrode The transmitter converts the measured voltage into the corresponding pH value using the Nernst equation Measuring system A...

Page 3: ...operating hours under extreme measuring conditions Once the sensor has been connected the sensor data are transferred automatically to the transmitter and used to calculate the current measured value...

Page 4: ...nce converter are a thing of the past EMC safety is guaranteed by screening measures for the digital transmission of measured values Intrinsically safe electronics mean operation in hazardous areas is...

Page 5: ...ata apply only if installing with Endress Hauser assemblies Make sure to follow the installation instructions in the Operating Instructions of the used assembly 15 15 A B A0024316 2 Electrode installa...

Page 6: ...rotective goggles and suitable gloves Pressure temperature ratings absolute A C 10 12 14 16 18 B 50 100 200 250 275 300 5 0 40 80 120 160 200 246 psi p bar T F 40 0 20 60 80 100 0 4 6 8 2 T C 120 135...

Page 7: ...Ag AgCl reference lead reference 4 Advanced Gel electrolyte 5 Ag AgCl reference lead pH 6 PTFE diaphragm 7 pH glass membrane 8 Temperature sensor NTC30K 3 4 5 6 1 2 7 8 9 12 0 47 120 225 360 4 72 8 8...

Page 8: ...llar 3 Ag AgCl reference lead reference 4 Advanced Gel electrolyte 5 Ag AgCl reference lead pH 6 PTFE diaphragm 7 pH glass membrane 8 Temperature sensor Pt100 Weight 0 1 kg 0 2 lbs Materials Electrode...

Page 9: ...FM Class I Div 2 in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters Biocompatibility Cytotoxicity verified in accordance with USP 2009 chapter 88 USP Class VI for diaphragm T V cert...

Page 10: ...nsors in tanks and pipes Product Configurator on the product page www endress com cpa450 Technical Information TI00183C Cleanfit CPA471 Compact stainless steel retractable assembly for installation in...

Page 11: ...rsion assembly for water and wastewater Modular assembly system for sensors in open basins channels and tanks Product Configurator on the product page www endress com cya112 Technical Information TI00...

Page 12: ...www addresses endress com...

Reviews: