CPS11D communication and data processing
Communication with the transmitter
Always connect digital sensors with Memosens technology to a transmitter with Memosens
technology. Data transmission to a transmitter for analog sensors is not possible.
Digital sensors can store measuring system data in the sensor. These include the following:
• Manufacturer data
– Serial number
– Order code
– Date of manufacture
• Calibration data
– Calibration date
– Slope at 25 °C (77 °F)
– Zero point at 25 °C (77 °F)
– Temperature offset
– Number of calibrations
– Serial number of the transmitter used to perform the last calibration
• Operating data
– Temperature application range
– pH application range
– Date of initial commissioning
– Maximum temperature value
– Hours of operation under extreme conditions
– Number of sterilizations
– Resistance of glass membrane
You can display the abovementioned data using the Liquiline CM44x, CM42 and Memobase Plus
CYZ71D.
Dependability
Reliability
Easy handling
Sensors with Memosens technology have an integrated electronics unit that stores calibration data
and other information (e.g. total operating hours and operating hours under extreme measuring
conditions). Once the sensor has been connected, the sensor data are transferred automatically to
the transmitter and used to calculate the current measured value. As the calibration data are stored
in the sensor, the sensor can be calibrated and adjusted independently of the measuring point. The
result:
• Easy calibration in the measuring lab under optimum external conditions increases the quality of
the calibration.
• Pre-calibrated sensors can be replaced quickly and easily, resulting in a dramatic increase in the
availability of the measuring point .
• Maintenance intervals can be defined based on all stored sensor load and calibration data and
predictive maintenance is possible.
• The sensor history can be documented using external storage media and evaluation programs, e.g.
Memobase Plus CYZ71D. Thus, the current application of the sensors can be made to depend on
their previous history.
Integrity
Data security thanks to digital data transmission
Memosens technology digitizes the measured values in the sensor and transmits the data to the
transmitter using a non-contact connection that is free from potential interference. The result:
• Automatic error message if sensor fails or connection between sensor and transmitter is
interrupted
• Immediate error detection increases measuring point availability