Installation
Installation instructions
Do not install the electrodes upside down. The angle of inclination must be at least 15° from the
horizontal. A smaller inclination angle is not permitted as it could cause an air bubble to form in the
glass sphere and prevent the inner electrolyte from completely wetting the pH diaphragm.
NOTICE
Before screwing in the electrode, make sure the threaded connection of the assembly is clean
and runs smoothly.
‣
Screw in the electrode finger-tight (3 Nm)! (Data apply only if installing with Hauser
assemblies.)
‣
Make sure to follow the installation instructions in the Operating Instructions of the used
assembly.
15°
15°
A
B
A0024316
2
Electrode installation; installation angle at least 15° from the horizontal
A
Permitted orientation
B
Forbidden orientation
Environment
Ambient temperature
NOTICE
Risk of damage due to frost
‣
The sensor must not be used if the temperature drops below –15 °C (5 °F).
Storage temperature
0 to 50 °C (32 to 122 °F)
Degree of protection
IP 68:
Memosens plug-in head, (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)
IP 68:
TOP68 plug-in head, autoclavable up to 135 °C (275 °F), (1 m (3.3 ft) water column,
50 °C (122 °F), 168 h)
IP 67:
GSA plug-in head (with closed connector system)