Installation

Installation instructions

Do not install the electrodes upside down. The angle of inclination must be at least 15° from the

horizontal. A smaller inclination angle is not permitted as it could cause an air bubble to form in the

glass sphere and prevent the inner electrolyte from completely wetting the pH diaphragm.

NOTICE

Before screwing in the electrode, make sure the threaded connection of the assembly is clean

and runs smoothly.

Screw in the electrode finger-tight (3 Nm)! (Data apply only if installing with Hauser

assemblies.)

Make sure to follow the installation instructions in the Operating Instructions of the used

assembly.

15°

15°

A

B

  A0024316

 2

Electrode installation; installation angle at least 15° from the horizontal

A

Permitted orientation

B

Forbidden orientation

Environment

Ambient temperature

NOTICE

Risk of damage due to frost

The sensor must not be used if the temperature drops below –15 °C (5 °F).

Storage temperature

0 to 50 °C (32 to 122 °F)

Degree of protection

IP 68:

Memosens plug-in head, (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)

IP 68:

TOP68 plug-in head, autoclavable up to 135 °C (275 °F), (1 m (3.3 ft) water column,

50 °C (122 °F), 168 h)

IP 67:

GSA plug-in head (with closed connector system)

Summary of Contents for Orbisint CPS11

Page 1: ...PTFE ring junction Can be used at pressures up to 17 bar abs 246 psi Process glass also for very alkaline applications BA and BT versions Process glass for applications in media containing hydrofluori...

Page 2: ...tegrated Ag AgCl reference system serves as the required reference electrode The transmitter converts the measured voltage into the corresponding pH value using the Nernst equation Measuring system A...

Page 3: ...operating hours under extreme measuring conditions Once the sensor has been connected the sensor data are transferred automatically to the transmitter and used to calculate the current measured value...

Page 4: ...nce converter are a thing of the past EMC safety is guaranteed by screening measures for the digital transmission of measured values Intrinsically safe electronics mean operation in hazardous areas is...

Page 5: ...ata apply only if installing with Endress Hauser assemblies Make sure to follow the installation instructions in the Operating Instructions of the used assembly 15 15 A B A0024316 2 Electrode installa...

Page 6: ...rotective goggles and suitable gloves Pressure temperature ratings absolute A C 10 12 14 16 18 B 50 100 200 250 275 300 5 0 40 80 120 160 200 246 psi p bar T F 40 0 20 60 80 100 0 4 6 8 2 T C 120 135...

Page 7: ...Ag AgCl reference lead reference 4 Advanced Gel electrolyte 5 Ag AgCl reference lead pH 6 PTFE diaphragm 7 pH glass membrane 8 Temperature sensor NTC30K 3 4 5 6 1 2 7 8 9 12 0 47 120 225 360 4 72 8 8...

Page 8: ...llar 3 Ag AgCl reference lead reference 4 Advanced Gel electrolyte 5 Ag AgCl reference lead pH 6 PTFE diaphragm 7 pH glass membrane 8 Temperature sensor Pt100 Weight 0 1 kg 0 2 lbs Materials Electrode...

Page 9: ...FM Class I Div 2 in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters Biocompatibility Cytotoxicity verified in accordance with USP 2009 chapter 88 USP Class VI for diaphragm T V cert...

Page 10: ...nsors in tanks and pipes Product Configurator on the product page www endress com cpa450 Technical Information TI00183C Cleanfit CPA471 Compact stainless steel retractable assembly for installation in...

Page 11: ...rsion assembly for water and wastewater Modular assembly system for sensors in open basins channels and tanks Product Configurator on the product page www endress com cya112 Technical Information TI00...

Page 12: ...www addresses endress com...

Reviews: