Orbisint CPS11D and CPS11

6

Hauser

Process

Process temperature

Version AA, AS:

-15 to 80 °C (5 to 176 °F)

Version BA, BT:

0 to 135 °C (32 to 275 °F)

Version FA:

0 to 70 °C (32 to 158 °F)

Process pressure (absolute)

Version AA, AS, FA:

1 to 7 bar (15 to 101 psi)

Version BA, BT:

1 to 17 bar (15 to 246 psi)

L

CAUTION

Pressurization of sensor due to prolonged use under increased process pressure

Risk of injury from glass breakage

Avoid excessive heating of such sensors if using them under reduced process pressure or under

atmospheric pressure.

When handling such sensors, wear protective goggles and suitable gloves.

Pressure-temperature

ratings (absolute)

A

C

10

12

14

16

18

B

50

100

200

250

275

300

5

0

40

80

120

160

200

246

[psi] p [bar]

T [°F]

40

0

20

60

80

100

0

4

6

8

2

T [°C]

120 135

-15

7

1

15

17

101

D

32

176

158

70

  A0025761

 3

Pressure-temperature ratings

A

Version AA, AS

B

Version BA, BT

C

Version FA

D

Atmospheric pressure

Minimum conductivity

Version AA, BA, BT, FA:

Min. 50 μS/cm (minimum flow; pressure and temperature must be

stable)

Version AS:

Min. 0.1 μS/cm (stainless steel flow assembly with grounding;

stable and minimum flow; pressure and temperature must be

stable)

pH range

Version AA, AS:

1 to 12 pH

Version BA, BT:

0 to 14 pH

Version FA:

0 to 10 pH

NOTICE

Risk of damage to electrode

Never use the electrode outside of the listed specifications!

Summary of Contents for Orbisint CPS11

Page 1: ...PTFE ring junction Can be used at pressures up to 17 bar abs 246 psi Process glass also for very alkaline applications BA and BT versions Process glass for applications in media containing hydrofluori...

Page 2: ...tegrated Ag AgCl reference system serves as the required reference electrode The transmitter converts the measured voltage into the corresponding pH value using the Nernst equation Measuring system A...

Page 3: ...operating hours under extreme measuring conditions Once the sensor has been connected the sensor data are transferred automatically to the transmitter and used to calculate the current measured value...

Page 4: ...nce converter are a thing of the past EMC safety is guaranteed by screening measures for the digital transmission of measured values Intrinsically safe electronics mean operation in hazardous areas is...

Page 5: ...ata apply only if installing with Endress Hauser assemblies Make sure to follow the installation instructions in the Operating Instructions of the used assembly 15 15 A B A0024316 2 Electrode installa...

Page 6: ...rotective goggles and suitable gloves Pressure temperature ratings absolute A C 10 12 14 16 18 B 50 100 200 250 275 300 5 0 40 80 120 160 200 246 psi p bar T F 40 0 20 60 80 100 0 4 6 8 2 T C 120 135...

Page 7: ...Ag AgCl reference lead reference 4 Advanced Gel electrolyte 5 Ag AgCl reference lead pH 6 PTFE diaphragm 7 pH glass membrane 8 Temperature sensor NTC30K 3 4 5 6 1 2 7 8 9 12 0 47 120 225 360 4 72 8 8...

Page 8: ...llar 3 Ag AgCl reference lead reference 4 Advanced Gel electrolyte 5 Ag AgCl reference lead pH 6 PTFE diaphragm 7 pH glass membrane 8 Temperature sensor Pt100 Weight 0 1 kg 0 2 lbs Materials Electrode...

Page 9: ...FM Class I Div 2 in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters Biocompatibility Cytotoxicity verified in accordance with USP 2009 chapter 88 USP Class VI for diaphragm T V cert...

Page 10: ...nsors in tanks and pipes Product Configurator on the product page www endress com cpa450 Technical Information TI00183C Cleanfit CPA471 Compact stainless steel retractable assembly for installation in...

Page 11: ...rsion assembly for water and wastewater Modular assembly system for sensors in open basins channels and tanks Product Configurator on the product page www endress com cya112 Technical Information TI00...

Page 12: ...www addresses endress com...

Reviews: