44
Nivotester FTR525
Hauser
Operation
9
Operation
9.1
Reading off the device locking status
The locking status of the device can be read as follows:
• Write protection via key lock (→ 42)
- The locking symbol lights permanently on the display.
- The note "
Key locked
" occurs if pressing a key.
• Write protection by changing the user
(→ 42)
- In the function
3121
the current user status is displayed.
- The locking symbol occurs when selecting a function without access right.
9.2
Reading off measured values
The measured values can be read from the display depending on the display mode
(→ 37).
In the functions
2510
"
Min/Max limit (1M)
" and
2520
"
Min/Max motion (2M)
", the
last minimum and maximum values can be read off as well.
9�3
Analysis
The following functions can be used to analyse the settings of the measuring system,
consisting of the Soliwave microwave barrier and the Nivotester FTR525.
9�3�1 Signal strength limit (1M)
Application
2000
→
Application analysis
2500
→
Signal strength limit (1M)
2540
This function is used to display the current signal strength of the limit detection (1M)
depending on the performed setup.
Example:
033
1M
2
540
2
1
3
4
5
Max. detection range
Set detection range
033
1M
2
540
27 Display signal strength of limit detection (1M)
1 Set lower limit of adjustment (LL) in function 2112 → 28
2 Set detection range (= 0-100% of signal output)
3 Percentage value of the current signal strength
4 Set upper limit of adjustment (UL) in function 2113 → 29
5 Current signal strength