Diagnostics and troubleshooting
Liquisys M CLM223/253
74
Hauser
E150
Distance between temp.
values in
value table too
small or not
monotonously increasing
Enter correct values in
value table
(minimum distance between
temperature values of 1 K required)
Yes
No
No
44
80
E151
USP/EP error
Yes
No
No
44
80
E152
PCS alarm
Check sensor and connection.
Yes
No
No
44
44
E153
USP/EP temperature error
Yes
No
No
80
44
E154
Below lower alarm
threshold for period
exceeding alarm delay
Perform manual comparison
measurement if necessary. Service
sensor and recalibrate.
Yes
No
No
-
3)
-
E155
Above upper alarm
threshold for period
exceeding alarm delay
Yes
No
No
-
-
E156
Actual value undershoots
alarm threshold (CC
setpoint) for longer than
the set permissible
maximum period
Yes
No
No
-
-
E157
Actual value exceeds
alarm threshold (CC
setpoint) for longer than
the set permissible
maximum period
Yes
No
No
-
-
E162
Dosage stop
Check settings in the CURRENT INPUT
and CHECK function groups.
Yes
No
No
-
-
E171
Flow in main stream too
low or zero
Restore flow.
Yes
No
No
-
-
E172
Switch-off limit for
current input exceeded
Check process variables at sending
measuring instrument. Change range
assignment if necessary.
Yes
No
No
-
-
E173
Current input < 4 mA
Check process variables at sending
measuring instrument.
Yes
No
No
-
-
E174
Current input > 20 mA
Check process variables at sending
measuring instrument. Change range
assignment if necessary.
Yes
No
No
-
-
1)
PV = Process variable
2)
If this error occurs, there is no possibility of starting a cleaning session (field F8 not applicable with this error).
3)
Current error messages not applicable via PROFIBUS
Error
no.
Display
Tests and/or remedial measures
Alarm contact
Error current
Autom.
cleaning start
PROFIBUS
status
Facty
User
Facty
User
Facty
User
PV
1)
Temp
Summary of Contents for Liquisys M CLM223/253
Page 112: ...www addresses endress com 71212629...