Commissioning
Liquisys M CLM223/253
44
Hauser
Basic version does not include functions in
italic
.
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
O
CURRENT OUTPUT
function group
Configuration of the current output (does
not apply for PROFIBUS).
O1
Select current output
Out1
Out 2
A characteristic can be selected for every
output.
O2
Select measured
variable for 2nd
current output
°C
mS/cm, M
, %
Contr
R237/R 266 = curr (current output 2) can
only be selected if O2 = Contr is selected
(relay board required).
O3 (1)
Enter or output
linear characteristic
Lin = linear
(1)
Sim = simulation (2)
Tab = table (3)
The characteristic can have a positive or
negative slope for the measured value
output.
In the case of actuating variable output
(O2 = Contr), an increasing current
corresponds to an increasing actuating
variable.
O311 Select current range
4 to 20 mA
0 to 20 mA
O312
0/4 mA value:
Enter corresponding
measured value
cond/ind:
0.00
μ
S/cm
MOhm:
0.00 k
cm
Conc:
0.00 %
Temp:
0.00 °C
Here you can enter the measured value at
which the min. current value (0/4 mA) is
applied at the transmitter output.
(Spreading: see Technical data.)
O313
20 mA value:
Enter corresponding
measured value
cond/ind:
2000 mS/cm
MOhm:
500 k
cm
Conc:
99.99 %
Temp:
150 °C
Here you can enter the measured value at
which the max. current value (20 mA) is
applied at the transmitter output.
(Spreading: see Technical data.)
O3 (2)
Simulate current
output
Lin = linear (1)
Sim = simulation
(2)
Tab = table (3)
Simulation is not ended until (1) or (3) is
selected.
For further characteristics, see O3 (1),
O3(3).
O321 Enter simulation
value
Current value
0.00 to 22.00 mA
Entering a current value results in this
value being directly output at the current
output.
O
OUTPUT
O1
Sel
.
Out
Out1
O2
Parameter
°C
O3
Sel
.
Type
lin
O311
Sel
.
Range
4
-
20
µS/cm
O312
0/4 mA
0.00
mS/cm
O313
20 mA
2000
O3
Sel
.
Type
sim
O321
mA
Simulat.
4.00