Liquisys M CLM223/253
Commissioning
Hauser
65
Coding
Field
Selection or range
(factory settings bold)
Display
Info
K
Function group
CONCENTRATION
Four different concentration fields can be
entered in this function group.
K1
Selection of
concentration curve,
to be used to
calculate the display
value
1
1 to 4
The curves are independent of each other.
Therefore, four different curves can be
defined.
K2
Selection of table to
be edited
1
1 to 4
When editing a curve, another curve
should be used to calculate the
corresponding values.
Example: When editing curve 2, either
curve 1, curve 3 or curve 4 should be active
(see K1).
K3
Select table option
read
edit
This selection applies to all concentration
curves.
K4
Enter number of
reference triplets
1
1 to 10
Each triplet consists of three numeric
values.
K5
Select triplet
1
1 to number of triplets in K4
assign
Any triplet can be edited.
If "assign", go to K9.
K6
Enter
uncompensated
conductivity
0.0 mS/cm
0.0 to 9999 mS/cm
The function chain K5 to K8 will run
through automatically as many times as
corresponds to the value in K4. Then the
system jumps to K9.
K7
Enter concentration
value for K6
0.00 %
0.00 to 99.99 %
Measuring unit selected as in A2.
Format selected as in A3.
K8
Enter temperature
value for K6
0.0 °C
-35.0 to 250.0 °C
K9
Message whether or
not the table status is
ok
yes
no
Only display
If not, then set table correctly (all previous
settings are kept) or back to measurement
mode (this makes the table invalid).
K
CONCENTRA
K1
act.curve
1
K2
editCurve
1
K3
Table
read
K4
No. Elem.
1
K5
Sel.Elem.
1
K6
mS/cm
conduct.
0.0
K7
%
concentr.
0.00
K8
C
°
Temp.val.
0.0
K9
Status ok
yes