Liquisys M CLM223/253
Maintenance
Hauser
83
Conductivity
For conductivity, the values in the following table are valid, if the cell constant k is set to the
nominal value according to column 2.
Otherwise: Display conductivity [mS/cm] = k[cm
-1
]
1 / R[k
]
The M
measurement is normally used for pure and ultrapure water and therefore is
only wise for cell constants where k= 0.01 or k = 0.1 cm
-1
.three-line system.
Resistance R
Cell constant k
Display for conductivity
Display for M
10
1 cm
-1
100 mS/cm
10 cm
-1
1000 mS/cm
100
0.1 cm
-1
1 mS/cm
1 k
cm
1 cm
-1
10 mS/cm
10 cm
-1
100 mS/cm
1000
0.1 cm
-1
0.1 mS/cm
10 k
cm
1 cm
-1
1 mS/cm
10 cm
-1
10 mS/cm
10 k
0.01 cm
-1
1
μ
S/cm
1 M
cm
0.1 cm
-1
10
μ
S/cm
100 k
cm
1 cm
-1
100
μ
S/cm
10 cm
-1
1 mS/cm
100 k
0.01 cm
-1
0.1
μ
S/cm
10 M
cm
0.1 cm
-1
1
μ
S/cm
1 M
cm
1 cm
-1
10
μ
S/cm
1 M
0.01 cm
-1
0.01
μ
S/cm
100 M
cm
0.1 cm
-1
0.1
μ
S/cm
10 M
cm
1 cm
-1
1
μ
S/cm
10 M
0.01 cm
-1
0.001
μ
S/cm
0.1 cm
-1
0.01
μ
S/cm
100 M
cm