Commissioning
Liquisys M CLM223/253
34
Hauser
For inductive sensors you must perform an airset before installation of the sensor, refer
to the chapter "Calibration"
25. In O1, select your output, e.g. "Out1" = output 1.
Press
F
to confirm.
Out1
Out2
26. In O3, select the linear characteristic.
Press
F
to confirm.
lin = linear (1)
sim = simulation
Tab = table
27. In O311, select the current range for your output,
e.g. 4 to 20 mA.
Press
F
to confirm.
4 to 20 mA
0 to 20 mA
28. In O312, enter the conductivity corresponding to
the minimum current value at the transmitter
output, e.g. 0
μ
S/cm.
Press
F
to confirm.
cond/ind:
0.00
μ
S/cm
MOhm:
0.00 k
cm
Conc:
0.00 %
Temp:
0.00 °C
29. In O313, enter the conductivity corresponding to
the maximum current value at the transmitter
output, e.g. 2000 mS/cm.
Press
F
to confirm.
The display returns to the initial display of the
"Current output" function group.
cond/ind:
2000 mS/
cm
MOhm:
500 k
cm
Conc:
99.99 %
Temp:
150 °C
30. Press
P
simultaneously to switch to the
measurement mode.
User input
Setting range
(Factory settings,
bold)
Display
O1
Sel
.
Out
Out1
O3
Sel
.
Type
lin
O311
Sel
.
Range
4
-
20
µS/cm
O312
0/4 mA
0.00
mS/cm
O313
20 mA
2000