Maintenance
Liquiline System CAT860
30
Hauser
Collecting vessel and hoses between the solenoid valve and collecting vessel
The automatic cleaning function of the Liquiline System CAT860 comprises the filter and
practically all the hoses. The collecting vessel and the hose between the solenoid valve and
collecting vessel do not fall within the scope of the automatic cleaning function. This excludes
the possibility of the measurement result being affected by the cleaning solution. Therefore, the
collecting vessel and the hose between the solenoid valve and collecting vessel must be cleaned
manually on a regular basis. The cleaning interval in a typical application is 1 week.
11.2
Replacing the pump hose and pump head
1. Open the sample preparation system door.
2. Press operating key 1 to activate on-site mode.
The status LED beside key 1 first starts flashing and is then lit continuously once the
on-site mode has been enabled. This may take a few minutes, see "Operation" section.
3. Remove the filter from the medium.
4. Press key 2.
The status LED beside key 2 comes on and the peristaltic pump rotates forward. The
medium in the hoses is now replaced by intake air.
5. Wait until all the hoses have been completely emptied.
6. Press key 2 again.
The pump stops and the status LED goes off.
7. Open the bayonet lock on the peristaltic pump.
8. Replace the hose and, where necessary, the pump head.
9. Close the bayonet lock on the peristaltic pump.
10. Check that all the hoses and connectors are properly seated.
11. Press operating key 1 to return to automatic mode.
The settings are accepted, the status LED beside operating key 1 goes out.
12. Close the sample preparation system door.