Installation
Liquiline System CAT860
20
Hauser
5.4
External compressed air connection
CAUTION
!
If not connected properly, this can result in injury and damage to the device
‣
Set a pressure of max. 5 bar (73 psi) at the pressure-reducing valve.
a0023277
Fig. 11: Connecting external compressed air
1
Hose, pump to analyzer
2
Hose, filter to pump
3
Hose, compressed air / cleaner
Prerequisites:
• Compressed air of 2 to 5 bar (29 to 73 psi)
• The compressed air must be filtered (40
μ
m) and free from water and oil
• No continuous air consumption
• Minimum nominal diameter for air ducts: 4mm (0.16 ")
1. Connect the compressed air line to the connection provided on the bottom of the housing.
2. Run the valve's rinsing air connection at an air pressure of 2 to 5 bar (29 to 73 psi).
Connect the pressure to the pressure-reducing valve.
5.5
Mounting the solenoid valve
The scope of delivery of the Liquiline System CAT860 includes a solenoid valve and a liquid
detector for mounting in the Liquiline System CA80. For further information on mounting
the valve in the analyzer, please refer to the enclosed kit instructions.