Liquiline System CAT860
Electrical connection
Hauser
21
5.6
Post-installation check
• Once installation is complete, check the sample preparation system and hoses for damage.
• After mounting, check whether all the screws are tightened correctly.
• When mounted, check all the connections to ensure that they are properly seated and
leak-proof.
• Make sure that the hoses cannot be removed without use of force.
6
Electrical connection
WARNING
!
The device is live!
Incorrect wiring can result in injury or fatality
‣
The electrical connection must only be carried out by a certified electrician.
‣
Technical personnel must have read and understood the instructions in this manual and
must adhere to them.
‣
Prior
to beginning any wiring work, make sure voltage is not applied to any of the cables.
NOTICE
The device does not have a power switch
‣
As soon as there is a power supply to the device, it starts up.
‣
The customer must provide a fuse with a maximum rating of 6.0 A at the installation
location. Please observe local installation regulations.
‣
The protective ground connection must be established before any other connections are
made. Disconnecting the protective ground could result in danger.
NOTICE
Torn off or buckled hoses can damage the device
‣
When folding down the carrier board make sure this does not damage any hoses.
Summary of Contents for Liquiline System CAT860
Page 2: ......
Page 41: ...Liquiline System CAT860 Endress Hauser W Warnings 4 Weight 38 Wiring 21 ...
Page 42: ...Liquiline System CAT860 42 ...
Page 43: ......