Technical data
Proline Promass 80
86
Hauser
Zero point stability
Example for max. measured error
A0013426
Abb. 41:
Max. measured error in % o.r. (example: Promass I, DN 25)
Flow values (example)
DN
Zero point stability
[mm]
[inch]
[kg/h] or [l/h]
[lb/min]
8
3/8"
0.150
0.0055
15
½"
0.488
0.0179
15 FB
½" FB
1.350
0.0496
25
1"
1.350
0.0496
25 FB
1" FB
3.375
0.124
40
1 ½"
3.375
0.124
40 FB
1 ½" FB
5.250
0.193
50
2"
5.250
0.193
50 FB
2" FB
13.50
0.496
80
3"
13.50
0.496
FB = Full bore (voller Nennweitenquerschnitt)
0
±0.5
±0.2
±1.0
[%]
0
1
2
4
6
8
10
12
14
16
18
t/h
Turn down
Flow
Maximum measured error
[kg/h] or [l/h]
[lb/min]
[% o.r.]
250 : 1
72
2.646
1.875
100 : 1
180
6.615
0.750
25 : 1
720
26.46
0.188
10 : 1
1800
66.15
0.100
2 : 1
9000
330.75
0.100