Technical data
Proline Promass 80
76
Hauser
10.1.6
Performance characteristics
Reference operating
conditions
• Error limits following ISO/DIN 11631
• Water, typ20 to +30 °C (+68 to +86 °F); 2 to 4 bar (30 to 60 psi)
• Data according to calibration protocol ±5 °C (±9 °F) and ±2 bar (±30 psi)
• Accuracy based on accredited calibration rigs according to ISO 17025
Performance characteristic
Promass A
Maximum measured error
The following values refer to the pulse/frequency output.
The additional measured error at the current output is typically ±5
μ
A.
Design fundamentals
77.
o.r. = of reading; 1 g/cc = 1 kg/l; T = medium temperature
• Mass flow and volume flow (liquid): ±0.15% o.r.
• Mass flow (gas): ±0.50% o.r.
• Density (liquid)
±0.0005 g/cc (under reference conditions)
±0.0005 g/cc (after field density calibration under process conditions)
±0.002 g/cc (after special density calibration)
±0.02 g/cc (over the entire measuring range of the sensor)
• Special density calibration (optional):
– Calibration range: 0.0 to 1.8 g/cc, +5 to +80 °C (+41 to +176 °F)
– Operation range: 0.0 to 5.0 g/cc, –50 to +200 °C (–58 to +392 °F)
• Temperature: ±0.5 °C ± 0.005 · T °C; (±1 °F ± 0.003 · (T - 32) °F)
Zero point stability
Example for max. measured error
A0013422
Fig. 37:
Max. measured error in % o.r. (example: Promass A, DN 2)
DN
Max. full scale value
Zero point stability
[mm]
[inch]
[kg/h] or [l/h]
[lb/min]
[kg/h] or [l/h]
[lb/min]
1
1/24"
20
0.73
0.0010
0.000036
2
1/12"
100
3.70
0.0050
0.00018
4
1/8"
450
16.5
0.0225
0.0008
0
±0.5
±0.1
[%]
0
50
100
kg/h
±0.2
±0.3
±0.4
10
5
20
30
40
60
70
80
90