Diagnostics and troubleshooting
Micropilot FMR60B HART
42
Hauser
X X X X
X X X
X
X
S801
4
5
3
6
1
S
2
X X X X
X X X
X
X
A0038013
1
Status signal
2
Event text
3
Status symbol
4
Diagnostic event
5
Status signal
6
Event number
If two or more diagnostic events occur simultaneously, only the diagnostic message with
the highest priority is shown.
"Active diagnostics" parameter
Key
Opens the message about the remedial measures.
Key
Acknowledge warnings.
Key
Back to operating menu.
11.4 Diagnostic event in the operating tool
If there is a diagnostic event in the device, the status signal appears on the top left in the
status section of the operating tool together with the associated icon for event behavior
according to NAMUR NE 107:
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
Click the status signal to see the detailed status signal.
The diagnostic events and remedial measures can be printed in the
Diagnostic list
submenu.
11.5 Adapting the diagnostic information
The event level can be configured:
Navigation: Diagnostics → Diagnostic settings → Configuration
11.6 Queued diagnostic messages
The display alternates between the queued diagnostic messages and the measured value.
Queued diagnostic messages can also be displayed in the
Active diagnostics
parameter.
Navigation: Diagnostics → Active diagnostics
Summary of Contents for Hart Micropilot FMR60B
Page 81: ...Micropilot FMR60B HART Index Endress Hauser 81 Write access 30 ...
Page 82: ......