Micropilot FMR60B HART
Mounting
Hauser
19
2.
1.
4
0.7 Nm
3.
4.
A0038224
1. If fitted: release the screw of the cover lock for the electronics compartment cover
using the Allen key.
2. Unscrew the electronics compartment cover from the transmitter housing and check
the cover seal.
3. Press the release mechanism and remove the display module.
4. Turn the display module to the desired position: maximum 4 × 90 ° in each direction.
Fit the display module on the electronics compartment in the desired position until it
clicks into place. Screw the electronics compartment cover back onto the transmitter
housing. If fitted: tighten the screw of the cover lock using the Allen key
0.7 Nm (0.52 lbf ft)±0.2 Nm (0.15 lbf ft).
5.3.7
Changing the installation position of the display module
The installation position of the display can be changed in the case of the dual
compartment housing, L-form.
A0048401