Micropilot FMR60B HART
Commissioning
Hauser
37
9.6
Device configuration
9.6.1
Level measurement in liquids
100%
0%
R
BD
D
L
F
E
A0016933
10
Configuration parameters for level measurement in liquids
R
Reference point of measurement
D
Distance
L
Level
E
"Empty calibration" parameter (= 0%)
F
"Full calibration" parameter (= 100%)
9.6.2
Commissioning with the Commissioning wizard
In FieldCare, DeviceCare, SmartBlue and on the display, the
Commissioning
wizard is
available to guide the user through the initial commissioning steps.
Complete this wizard to commission the device.
For each parameter, enter the appropriate value or select the appropriate option.
NOTE
If you exit the wizard before completing all required parameters, the changes you made
will be saved. For this reason, the device may be in an undefined state!
In this case, a reset to the default settings is recommended.
9.7
Recording the Echo curve
Record the current Echo curve as a reference echo curve for subsequent diagnostic
purposes
After the measurement has been configured, it is recommended to record the current Echo
curve as a reference echo curve.
The
Save reference curve
parameter in the
Echo curve
submenu is used to record the
Echo curve.
Diagnostics → Echo curve → Save reference curve
‣
Under the
Save reference curve
parameter, activate the
Customer reference curve
option
9.8
"Simulation" submenu
Simulation of a process variable, a pulse output or a diagnostic event.