The installation of an INS proximity switch in the assembly enables reliable detection of this
invalid operating status, thus triggering an alarm or causing the dosing process to be switched
off if necessary.
Below the minimum flow rate, the sensor current is more sensitive to flow fluctuations. For
abrasive media, it is recommended not to exceed the minimum flow. If suspended solids are
present, which may form deposits, the maximum flow rate is recommended.
Temperature
Changes in the temperature of the medium affect the measured value:
• Increases in temperature result in a higher measured value (approx. 4 % per K)
• Decreases in temperature result in a lower measured value.
Using the sensor in combination with the Liquisys CCM223/253 enables automatic
temperature compensation (ATC). Recalibration in the case of temperature changes is not
necessary.
1. If automatic temperature compensation is disabled at the transmitter, the temperature
must be maintained at a constant level following calibration.
2. Otherwise, recalibrate the sensor.