Electrical connection
CCS240/241
16
Hauser
6.1
Connecting the sensor
‣
Install the grounding bar (order number 51501086) in accordance with the accompanying
instructions in order to guarantee a high reading stability.
NOTICE
Measured errors due to faulty connection
‣
When connecting the sensor cable, make sure that the black semi-conductor layer is
removed as far as the inner shield.
The sensors have a fixed cable with a maximum length of 3 m (9.8 ft).
‣
Connect the sensors to the transmitter according to the following diagram:
Sensor: assignment
Sensor: core
Transmitter: terminal
Outer shield
S
Anode
[A] red
91
Cathode
[K] transparent
90
NTC temperature
sensor
Green
11
NTC temperature
sensor
Brown
12
GN
1
2
3
4
5
6
7
8
BN
A0036973
5
Structure of the sensor cable
1
Outer shield
2
Inner shield, anode
3
Semi-conductor layer
4
Inner insulation
5
Inner conductor, measured signal
6
Temperature sensor connection
7
2nd insulation
8
Outer insulation