Maintenance
CCS240/241
24
Hauser
9.2.2
Replacing the membrane
1
2
3
A0037110
9
Replacing the membrane
1
Measuring chamber
2
Membrane cap
3
Screw cap
1. Unscrew the measuring chamber (1).
2. Unscrew the front screw cap (3).
3. Remove the membrane cap (2) and replace it with a CCY14-WP replacement cartridge.
4. Refill the measuring chamber with electrolyte CCY14-F(→ 25).