Operation
CA76NA
42
Hauser
Group
name
Starting
address
Size
(bytes)
Format
Parameter name
Description
Unit
Channel 2
58
4
REAL
Na concentration C2
Sodium concentration of
channel 2
ppb,
µg/l
62
1
BYTE
Status C2
For assignment, see
"Channel status" table
63
1
BYTE
Reserve C2
64
2
INT16
Measuring time C2
Set measuring time in
automatic mode
min
Channel 3
66
4
REAL
Na concentration C3
Sodium concentration of
channel 3
ppb,
µg/l
70
1
BYTE
Status C3
Assignment of channel
status → 44
71
1
BYTE
Reserve C3
72
2
INT16
Measuring time C3
Set measuring time in
automatic mode
min
Channel 4
74
4
REAL
Na concentration C4
Sodium concentration of
channel 4
ppb,
µg/l
78
1
BYTE
Status C4
Assignment of channel
status → 44
79
1
BYTE
Reserve C4
80
2
INT16
Measuring time C4
Set measuring time in
automatic mode
min
Channel 5
82
4
REAL
Na concentration C5
Sodium concentration of
channel 5
ppb,
µg/l
86
1
BYTE
Status C5
Assignment of channel
status → 44
87
1
BYTE
Reserve C5
88
2
INT16
Measuring time C5
Set measuring time in
automatic mode
min
Channel 6
90
4
REAL
Na concentration C6
Sodium concentration of
channel 6
ppb,
µg/l
94
1
BYTE
Status C6
Assignment of channel
status → 44
95
1
BYTE
Reserve C6
96
2
INT16
Measuring time C6
Set measuring time in
automatic mode
min
1)
If the channel is active, the remaining measuring time is output; if no measuring time has been assigned:
-1; if device is 1-channel device: -2
2)
If the channel is active, the remaining measuring time is output; if no measuring time has been assigned:
-1
Output data (PROFIBUS to analyzer)
Group name
Starting
address
Size
(bytes)
Format
Parameter name
Description
Remote control
0
2
2 BYTE
Remote control
Assignment of remote control
→ 44
2
2
INT16
Calibration interval
Permitted values, index
calibration interval → 45
4
2
INT16
Number of
regenerations
Maximum number of
regenerations is limited,
maximum permitted number:
(calibration interval [h])/2)-1