Operation
CA76NA
36
Hauser
Maintenance
Function
Options
Info
pH Calibration
Displays measured values and entry fields when
calibrating the pH electrode.
pH electrode calibration
pH1 pH ---
Enter the pH value of the first buffer solution
used
The pH value of the first buffer solution
should be lower than that of the second
buffer solution.
pH2 pH ---
Enter the pH value of the second buffer solution
used
The pH value of the second buffer solution
should be greater than that of the first
buffer solution.
Temp.
Displays the zero point shift
S mV/D
Enter the average temperature of the buffer
solutions, deactivatable temperature
compensation
E0 mV
Displays the slope
Meas.pot. mV
Displays the measuring potential
Meas.value pH
Displays the measured value for pH
Reagent Exchange
‣
Run after replacing the standard solution.
Interface 20mA
Specify a 4-20 mA current signal for each
channel, e.g. for loop checks to the central
process control system.
1.
Enter the desired mA value.
2.
Switch to ON to activate the outputting of
a current signal for the specific channel.
3.
Press to confirm.
The outputting of a current signal is
automatically set to off when you
exit the menu item. The analyzer
outputs the current mA value.