Mounting
CA76NA
16
Hauser
ESC
!
190 (7.48)
20 (0.79)
450 (17.7)
424 (16.7)
850 (33.5)
817 (32.2)
13 (0.51)
(4x)10 x 7 (0.39 x 0.28)
31 (1.22)
61 (2.40)
91 (3.58)
121 (4.76)
151 (5.94)
181 (7.13)
217 (8.54)
A0047739
4
Analyzer CA76NA. Unit of measurement mm (in)
5.1.3
Installation site
Please note the following:
1. Protect the device against mechanical vibrations.
2. Protect the device against chemical exposure.
3. Do not expose the device to very dusty environments.
4. Install the device in a dry environment.
5. Make sure that the wall has sufficient load-bearing capacity and is fully
perpendicular.
6. Make sure that the device is horizontally aligned and is mounted on a vertical surface
(mounting plate or wall).
7. Protect the device from additional heating (e.g. from heaters or direct sunlight).
Comply with the following minimum spacing requirements:
• at least 10 mm (0.39 in) at the sides of the analyzer
• at least 550 mm (21.7 in) in front of the analyzer
• at least 200 mm (7.87 in) under the analyzer since cables and water conduits are
connected from below