2. Screw the bottle into the head provided. In doing so, make sure you do not touch the
hoses.
11
Connected bottle for sodium standard solution, incl. head
8.1.4
Setting the sample flow
The control valve is used to adjust the sampling volume so that the sample flows off evenly
at the overflow.
1.
12
Control valve
Set the sample flow to 5 to 10 l/h (1.32 to 2.64 gal/h) at the control valve.
2. Wait until sample flows off evenly via the overflow.
3. Repeat the process for all available channels.
8.2
Setting up PROFIBUS communication
1. In the main menu, select
Parameters
/
Outputs
/
Profibus
.
2. Configure the PROFIBUS slave address of the analyzer.
3. Switch off the device.
4. Connect the PROFIBUS cable to the PROFIBUS interface .