Electrical connection
CA76NA
20
Hauser
Mains voltage
Multi-range power unit for 100 to 240 V AC
The analyzer is fitted with a fuse, T 1.25 A, for the 215
to 240 V AC voltage level. If the analyzer is operated
with 100 to 130 V AC, replace the fuse with the T 2.5 A
fuse supplied. The fuse is located in the cover of the
electronics unit.
Analog outputs
• X12: current output, channel 1 + 2
• X15: current output, channel 3 + 4
• X23: current output, channel 5 + 6
Digital outputs
• X1: relay, 1 alarm
• Open contact on error:
COM-NO
• Closed contact on error:
COM-NC
• X3: relay 2, warning
• Open contact on error:
COM-NC
• Closed contact on error:
COM-NO
Terminal diagram with PROFIBUS
L1 N PE
A+
COM
COM
B+
COM2
NC2
NO2
NO1
COM1
NC1
X 100
X12A
X23
CH1
CH2
X3
X1
F01
PR
OF
IBU
S
X5
B
A
GND
SH
X12B
X7
X8
A0041292
L1
N
PE
NO1 CO
M1
NC1 NO2 CO
M2
NC2 A+
CO
M
B+
CO
M
B
A
GND SH
X100
Power supply
100 to 240 V AC,
50/60 Hz
X1
Relay 1
Alarm
X3
Relay 2
Warning
X12A
4 to 20
mA
Channel 1
X12B
4 to 20
mA
Channel 2
PROFIBUS cable (internal)
Mains voltage
Multi-range power unit for 100 to 240 V AC
Analog outputs
X12: current output, channel 1 + 2
Digital outputs
• X1: relay 1, alarms
• Open contact on error: COM-NO
• Closed contact on error: COM-NC
• X3: relay 2, warnings
• Open contact on error: COM-NC
• Closed contact on error: COM-NO
If the CA76NA is the last device in the bus segment, the two jumpers must be set to X7
and X8 on the PROFIBUS interface card to incorporate the terminating resistors. If the
analyzer is not the last device in the bus segment, the jumpers must be removed from X7
and X8 on the PROFIBUS interface card.