Operation
CA76NA
44
Hauser
Channel status
Bit 7
Bit 6
Channel status
Description
0
0
bad
pH value is too small (the pH value is smaller than the
set limit)
0
1
uncertain
All calibration errors (warnings), stock solution
empty, insufficient water flow
1
0
good
If no errors or warnings occurred during the
measurement
Remote control
Bit
Remote control
Description
0
Start calibration
Starts calibration procedure
1
Start meas. ch. 1
Starts measurement of channel 1
(no time limit, switched off by "off")
2
Start meas. ch. 2
Starts measurement of channel 2
(no time limit, switched off by "off")
3
Start meas. ch. 3
Starts measurement of channel 3
(no time limit, switched off by "off")
4
Start meas. ch. 4
Starts measurement of channel 4
(no time limit, switched off by "off")
5
Start meas. ch. 5
Starts measurement of channel 5
(no time limit, switched off by "off")
6
Start meas. ch. 6
Starts measurement of channel 6
(no time limit, switched off by "off")
7
Start regeneration
Starts the automatic regeneration of the Na electrode
8
Off
Stops the technique currently running, the device then
switches to standby
9
Start automatic
Starts the automatic program sequence
10
Set calibration interval
Sets the value for the calibration interval provided under
"Calibration interval" (byte 2 and 3)
11
Set number of regenerations
Sets the value for the number of regenerations provided under
"Number of regenerations" (byte 4 and 5)
Sample flow
Bit
Remote control
Description
0
-
-
1
No sample channel 1
Insufficient sample flow on channel 1
2
No sample channel 2
Insufficient sample flow on channel 2
3
No sample channel 3
Insufficient sample flow on channel 3
4
No sample channel 4
Insufficient sample flow on channel 4
5
No sample channel 5
Insufficient sample flow on channel 5
6
No sample channel 6
Insufficient sample flow on channel 6
7
-
-