Running, pausing, continuing and stopping programs
Biometra TAdvanced
74
Version 02_20
Fig. 58 Spreadsheet view
Fig. 59 Display screen during a program pause (graphical view)
Note
: Pressing the [P
AUSE
] button requires manual interaction. The Biometra
TAdvanced software also allows pauses to be programmed (see section "Editing steps"
p. 49), then the pause step is automatically started by the device.
Note
: If the device is switched off during a pause step, this will be recognized as a
power failure (a corresponding error message is written to the run log file). If, for
example, the last step in a program is a pause, the program has to be stopped by
pressing the [S
TOP
] button (see section "Stopping the program" p. 75), before
switching off the Biometra TAdvanced.
Biometra TAdvanced
Twin