Running, pausing, continuing and stopping programs
Biometra TAdvanced
72
Version 02_20
Fig. 55 Display during the run in a graphical, spreadsheet and gradient view
The following parameters are summarized in the headline in the table:
Program name
Sample block temperature
Estimated end
Heated lid temperature
Remaining runtime
Sequential number of working step
If the heated lid is set to preheating O
N
(see section "Preheating the heated lid" p. 48),
during the preheating of the heated lid the word "Preheat" will be displayed in the S
TEP
field in the table. After the heated lid has reached the target temperature and the 40s
equilibration phase has passed, the first program step starts. During the program run,
the current step is highlighted in yellow in the graphical and spreadsheet view. In the
graphical view, each step is divided into a heating phase and a hold time and both
parts are highlighted separately. Time and temperature increments and ramp rate
adjustments cannot be shown in the graphical view. Instead the symbols
T,
t or !
are displayed (see section "Programming modes" p. 44).
If it is a gradient program, G
RAD
is also displayed in brackets in the N
AME
field after the
program name (for example see Fig. 52).
During a gradient step, the temperatures present in each individual column in the
sample block can be checked in the gradient display.