Repair
Condumax CLS15D/16D/21D
22
Hauser
Recommended intervals for seal replacement (reference values)
Application
Window
Media with temperatures from 50 to 100 °C (122 to 212 °F)
Approx. 18 months
Media with temperatures < 50 °C (122 °F)
Approx. 36 months
Sterilization cycles, max. 150 °C (302 °F), 45 min.
Approx. 400 cycles
To ensure your sensor is operational again after being exposed to very high loads, you can
have it regenerated in the factory. In the factory, the sensor is fitted with new seals and
recalibrated.
Please contact your sales office for information on replacing the seal and recalibration in the
factory.
8.2
Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions at
www.endress.com/support/return-material
.
8.3
Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Observe the local regulations.