36
REPROCESSING
The instruments, CombiTorque, handpiece holder, and handpiece supplied with the Piezon 150 and
Piezon 250 must be cleaned and sterilized before the first use. Reusable accessories must be cleaned
and sterilized following each use. A listing of the reusable accessories is provided in the table below.
Diamond-coated instruments are indicated for single use only and must not be reused.
Follow present-day regulations enforced in the country about reprocessing.
Clean the exterior surfaces of the unit only with an alcohol-based, commercially available (ethanol,
isopropanol), colorless disinfectant. The use of scouring powder or an abrasive sponge will damage its
surface.
FB-569_HS_2014_10_Notice_PIEZON 150_250.indd 36
7/3/2015 1:53:20 PM