The end of the socket
or the welding flange must not
have an
extension into the tank in Fig. 6.
NON-METAL TANK
For level meter installation in non
-
conductive tanks it is neces-
sary to use a metal sheet (diameter
greate
r than 200 mm) be-
neath the process fitting when screwing it in. Make sure that
the plate has direct contact with the process fitting.
CONCTRETE SILO
For installation of the level meter on a concrete roof the diameter �
of the hole must be greater than the thick-
ness a of the concrete, see Fig. 8.
If the thickness a of the concrete is greater than the diameter �
of the hole, install the level meter in a recess,
see. Fig. 9.
LEVEL METER PROTECTION
The level meter must not be installed in places with direct solar
radiation and must be protected against weather effects.
If the installation in places with direct solar radiation is inevita-
ble, it is necessary to mount a
shielding cover
above the level
meter.
All types
Except
EMKOTDR-60-20
Fig. 8: Level meter installation on the
roof of the concrete silo
Fig. 9: Level meter installation on the
roof of the concrete silo
Fig. 7: Incorrect welding flange mounting to the tank
Fig. 6: Incorrect welding flange mounting to the tank
All types
Except
EMKOTDR-60-20
EMKOTDR-60-20,22,23
Fig. 10: Solar radiation shielding cover
All types