5. Installation and putting into operation
Please follow next 3 steps:
•
Installation
•
Electrical connection
•
Setting
6.
Installation instructions
BASIC INFORMATIONS
Install the level meter into the upper lid of the tank or reser-
voir using a welding flange or fastening
nut.
The min. distance to install the level meter into a lid or a ceil-
ing of a tank from the tank wall is given in table right
.
Otherwise, the level meter install as far as possible from the
walls, to the middle between the wall and the vertical inlet,
see Fig. 1. a
nd 2.
type of wall
d
(without ref. tube)
d
(with ref. tube)
metal
≥ 300 mm
any distance
non-metal
≥ 500 mm
any distance
All types