TB-6594
Page 12 of 12
© 2016 DESCO INDUSTRIES, INC.
Employee Owned
Calibration
Frequency of recalibration should be based on the
critical nature of those ESD sensitive items handled
and the risk of failure for the ESD protective equipment
and materials. In general, EMIT recommends that
calibration be performed annually.
Use the EMIT 50424 Limit Comparator and EMIT 50784
5-Pound Electrode to perform periodic testing (once
every 6-12 months) of the SmartLog Pro™. The Limit
Comparator can be used on the shop floor within a few
minutes virtually eliminating downtime, verifying that the
tester is operating within tolerances.
See
TB-6581
for more information.
Figure 22. EMIT 50424 Limit Comparator
Figure 22. EMIT 50784 5-Pound Electrode for Limit
Comparator
Specifications
SmartLog Pro™
Input Voltage and
Frequency
(External Adapter)
AC/DC Power Adapter
Power Input:
100-240VAC, 50/60 Hz
Power Output:
5VDC, 3.0A
Cable Length:
5 ft. (1.5 m)
Operating Temperature 70°F to 85°F (21°C to 30°C)
for 1 gigohm test limit
41°F to 85°F (5°C to 30°C)
for all other test limits
Environmental
Requirements
Indoor use only at altitudes
less than 6500 ft. (2 km)
Maximum relative humidity
of 80% up to 85°F (30°C)
decreasing linearly to 50% @
85°F (30°C)
Maximum relative humidity of
50% at 1 gigohm setting
Dimensions
6.75" x 5.00" x 1.75"
(17.1 cm x 12.7 cm x 4.4 cm)
Weight
1.8 lbs (0.8 kg)
Test Accuracy
±20% for 1 gigohm footwear
test limit
±10% for all other test limits
Test Switch Voltage
5 VDC @ open circuit
Wrist Strap and
Footwear Test Voltage
30 VDC @ open circuit
Test current is limited by
resistors and varies on
the test range setting (100
kilohms - 1 gigohm)
Dual Independent Foot Plate
Dimensions
14.0" x 16.0" x 0.9"
(35.6 cm x 40.1 cm x 2.3 cm)
Weight
7.5 lbs (3.4 kg)
Limited Warranty, Warranty Exclusions, Limit of
Liability and RMA Request Instructions
See the EMIT Warranty -
http://emit.descoindustries.com/Warranty.aspx