 Page 1 of 12

© 2016 DESCO INDUSTRIES, INC.

Employee Owned

SmartLog Pro™

Installation, Operation and Maintenance

Description

The patented* EMIT 50780 SmartLog Pro™ system is 

designed for fast, frequent, and accurate testing of ESD 

personnel grounding items.  Its unique design embeds 

an ESD tester, computer, touchscreen display, proximity 

badge reader and barcode scanner into a compact 

metal enclosure.
By touching the solid-state switch once, the SmartLog 

Pro™ independently tests the resistance path limits of 

the worn wrist strap and both worn ESD footwear in less 

than 2 seconds.  It may also test a worn ESD garment 

if it is used as part of personnel grounding path.  Test 

results are electronically stored in the SmartLog Pro™ 

and easily downloaded to a PC for logging records 

and evaluation.  This product can be used as one of 

the tools to fulfill the ANSI/ESD S20.20 section 7.3 

“Compliance Verification Plan.”
Paperless data can enhance operator accountability, 

immediately identifying problems while reducing logging 

and auditing costs.  There is no need to dedicate a 

computer for each test station.  The SmartLog Pro™ is 

a complete system including all required components.  

Operator identification can be accomplished using the 

touchscreen keypad, scanning a barcode badge or 

waving a proximity badge (verify compatibility with the 

factory).

May 2016

Figure 1.  EMIT 50780 SmartLog Pro™

The SmartLog Pro™ can test either single or dual-wire 

wrist straps and its split footplate design provides 

individual footwear testing all in a single test.  With 

the use of TEAM5 Software, the SmartLog Pro™ can 

be programmed to assign unique test requirements to 

personnel.  An individual can be required to test either 

wrist strap only, footwear only or a combination of the 

two.
If a resistance path is below or exceeds the set limits, 

the SmartLog Pro™ will indicate failure via audio and 

visual alarms.  Use the included relay terminal to 

promote access control for passed tests.
The SmartLog Pro™ can also be networked to a 

company’s Intranet using its embedded Ethernet 

module.  The SmartLog Pro™ is calibrated to NIST 

traceable standards.

ESD Association Information

“Compliance verification should be performed prior to 

each use (daily, shift change, etc.).  The accumulation 

of insulative materials may increase the foot grounder 

system resistance.  If foot grounders are worn outside 

the ESD protected area testing for functionality 

before reentry to the ESD protected area should be 

considered.”  ESD SP9.2 APPENDIX B - Foot Grounder 

Usage Guidance
“Process monitoring (measurements) shall be conducted 

in accordance with a Compliance Verification Plan that 

identifies the technical requirements to be verified, the 

measurements limits and the frequency at which those 

verifications shall occur...Compliance verification records 

shall be established and maintained to provide evidence 

of conformity to the technical requirements.
The test equipment selected shall be capable of 

making the measurements defined in the Compliance 

Verification Plan.” (ANSI/ESD S20.20) section 7.3 
ANSI/ESD S20.20 Table 1 Flooring-Footwear Systems 

Technical Requirements Recommended Range “less 

than 3.5 x 10

7

 ohms measured per ANSI/ESD STM 

97.1.”
“Typical test programs recommend that wrist straps that 

are used daily should be tested daily.  However, if the 

products that are being produced are of such value that 

knowledge of a continuous, reliable ground is needed, 

and then continuous monitoring should be considered or 

even required.”  (ESD Handbook ESD TR 20.20 section 

5.3.2.4.4)
Note: SmartLog Pro™ provides wrist strap test per 

IEC 61340-5-1 Clause A.1 and footwear testing per 

IEC 61340-5-1 Clause A.2 with upper limits < 3.5 x 10

7

 

ohms.

TECHNICAL BULLETIN TB-6594 

*US Patents 6,078,875 and 6,809,522

Made in the

United States of America

Summary of Contents for SmartLog Pro 50780

Page 1: ...r exceeds the set limits the SmartLog Pro will indicate failure via audio and visual alarms Use the included relay terminal to promote access control for passed tests The SmartLog Pro can also be netw...

Page 2: ...tem Overview 4 Item and Accessories 4 Packaging 5 Features and Components 6 Installation 6 Hardware Setup 7 10mm Wrist Cord Adapter 8 Relay Terminal 8 Network Setup 9 Test Limit Configuration 10 TEAM5...

Page 3: ...artLog Pro SmartLog Pro PC with TEAM5 Software Supervisor Department Manager Department Manager Department Manager Figure 2 Illustration of the SmartLog Pro System collecting and transferring ESD data...

Page 4: ...ator for Testers Use the SmartLog Pro along with the TEAM5 Software to automate the collection of employee ESD testing This software is offered in two tiers TEAM5 and TEAM5 Enterprise Click here to vi...

Page 5: ...TB 6594 Page 5 of 12 2016 DESCO INDUSTRIES INC Employee Owned Features and Components Figure 3 SmartLog Pro features and components A B C D E F G H I J K L M N O...

Page 6: ...the tinned wire termination of the included ground cord to this terminal Connect the ring terminal termination of the cord to equipment ground This connection will remove any static charge from the u...

Page 7: ...clear the clip at the base of the plunger s shaft when inserting into the jack 6 Connect the SmartLog Pro to the bracket Use the included thumbscrew to secure the SmartLog as shown below Figure 5 Secu...

Page 8: ...Ethernet communication module DYNAMIC IP PROCEDURE 1 Verify that the Ethernet cable is securely connected to your network and SmartLog Pro The LEDs on the Ethernet port will illuminate when a connect...

Page 9: ...quired to establish communication between the SmartLog Pro and the TEAM5 Server Software Figure 13 Locating the IP address in the Admin Menu Figure 14 Locating the Newtork button in the Admin Menu Fig...

Page 10: ...tion Software on the market When used with the EMIT SmartLog TEAM5 allows manufacturers and assemblers to automate the collection of employee ESD testing It has a set of robust employee management fun...

Page 11: ...if using proximity badges See Figure 20 2 Follow the prompt on the SmartLog s display 3 When performing a footwear test be sure to place both feet on the dual foot plate one foot per plate NOTE Keep t...

Page 12: ...Adapter Power Input 100 240VAC 50 60 Hz Power Output 5VDC 3 0A Cable Length 5 ft 1 5 m Operating Temperature 70 F to 85 F 21 C to 30 C for 1 gigohm test limit 41 F to 85 F 5 C to 30 C for all other te...

Page 13: ...EMIT Smart ESD Process Control Smart ESD Process Control 000 Checkout Products What s New Videos SmartLog V5 FAQ http emit descoindustries com 1 of 5 6 3 2016 10 45 25 AM...

Page 14: ...trol Monitors Ionization Testers Access Control Data Acquisition Measurement Factory Closeouts Shop from our other Desco Industries websites Europe United Kingdom Asia http emit descoindustries com 2...

Page 15: ...EMIT Smart ESD Process Control http emit descoindustries com 3 of 5 6 3 2016 10 45 25 AM...

Page 16: ...EMIT Smart ESD Process Control http emit descoindustries com 4 of 5 6 3 2016 10 45 25 AM...

Page 17: ...Policy Terms Conditions Our Factories Additional Resources Technical Information Resources Ask an ESD Question Find a Sales Rep or Reseller Reseller Registrations Site Map Sign up for email updates g...

Reviews: